LFQ, low input#

archived Quantification — DIA, Orbitrap Astral, low sample input

This module is archived

Superseded by the LFQ, precursor ion (Astral, low input) and LFQ, precursor ion (timsTOF, low input) modules, following this discussion. The module and its submission form remain open — existing and new benchmark runs stay comparable here — but for new low-input benchmarking we recommend using one of the modules above instead.

Compares the sensitivity and quantification accuracy of workflows for data acquired with data-independent acquisition (DIA) on an Astral at low sample input (single-cell-scale amounts).

For bulk sample analyses on the same instrument type, see LFQ, precursor ion (Astral).

Open in web app

At a glance#

Acquisition

DIA

Instrument

Orbitrap Astral, low input

Level

Precursor ion

Metric

Epsilon (quantification accuracy)

What this module tests#

Search-specific instructions, metric definitions, and result-column descriptions are shared with the bulk DIA Astral module; see that page for the full detail. This module is not designed to evaluate later-stage post-processing (e.g. missing value replacement or manual filtering).

Dataset#

A subset of the Astral DIA data described in Bubis et al., 2024, using only the series with 240:10 (condition “A”) and 200:50 (condition “B”) ratios, three replicates each. Samples are a mixture of HeLa (Thermo Scientific, Pierce™ HeLa Protein Digest Standard, 88328) and yeast (Promega, MS Compatible Yeast Protein Extract Digest, S. cerevisiae, V7461), combined in 0.1% TFA.

Download:

Important

Do not rename the downloaded raw files.

Run your workflow#

Run your analysis on the downloaded files, then follow Your First Submission to upload, inspect, and (if you’d like) submit your results publicly.

Tool-specific setup#

Table: input files required for metric calculation and public submission

Tool

Input file

Parameter file

AlphaDIA

precursors.parquet/.tsv (v2+) or precursor.matrix.tsv + precursors.tsv (v1)

log_alphadia.txt

Custom

custom_input.tsv

—

DIA-NN

report.tsv or report.parquet

report.log.txt

FragPipe

combined_ion.tsv

fragpipe.workflow

FragPipe (DIA-NN quant)

report.tsv or report.parquet

fragpipe.workflow

MSAID

MSAID_output.tsv

MSAID_params.csv

MaxQuant

evidence.txt

mqpar.xml

PEAKS

lfq.features.csv

*.txt

Spectronaut

*.tsv

ExperimentSetupOverview.txt

For per-tool setup steps, see the DIA Astral bulk module’s tool-specific setup — the same tools and general workflow apply here.

Custom format

If your tool isn’t listed above, upload a tab-delimited table with:

  • Sequence — unmodified peptide sequence

  • Proteins — ;-separated identifiers, including the species flag (e.g. _YEAST)

  • Charge — precursor charge

  • Modified sequence — sequence with localized modifications, in ProForma

  • one quantitative column per sample: 20231123_DIA_240k_20Th_40ms_FAIMSCV-48_gas3p8_240pg_10pg_H_Y_r1–r3 (condition A), ..._200pg_50pg_H_Y_r1–r3 (condition B)

The table must not contain non-validated ions.

Parameters tracked for public submission#

Parameters tracked, result columns, and how the metric is calculated are the same as the DIA Astral bulk module; see that page for the full list.

Contact us or open an issue with any problems.