LFQ, precursor ion (ZenoTOF)#
beta Quantification — DIA, ZenoTOF 8600, precursor-ion level
This module is in beta stage
This module has been validated and is suitable for submissions. However, metrics or thresholds may be refined as we gather more community results and feedback.
Compares label-free quantification (LFQ) accuracy and sensitivity for workflows run on data-independent acquisition (DIA) data, namely Zeno SWATH DIA with 85 variable windows, on a ZenoTOF 8600 (SCIEX).
At a glance#
Acquisition |
DIA, Zeno SWATH (85 variable windows) |
Instrument |
ZenoTOF 8600 |
Level |
Precursor ion (modified sequence + charge) |
Metric |
Epsilon (quantification accuracy) |
What this module tests#
This module is best suited to evaluate the impact of, among others:
search engine identification
peak picking
low-level ion signal normalization
It is not designed to evaluate later-stage post-processing of quantitative data (e.g. missing value replacement or manual filtering); please upload data without that kind of post-processing.
Dataset#
A not-yet-released ZenoTOF 8600 DIA dataset using the same sample composition (conditions “A” and “B”) as described in Van Puyvelde et al., 2022: commercial peptide digest standards of Escherichia coli, yeast, and human, at logarithmic fold changes (log2FC) of 0, −1, and 2 respectively.
Peptides were separated on an IonOpticks Aurora® XS Ultimate column (25 cm × 75 µm, 1.7 µm) at 0.250 µL/min, 40 °C, over a 15-min gradient (3% B hold, ramp to 35% over 15 min, then 65% and 80% B, wash, re-equilibrate). An OptiFlow Pro Nano source with NanoCal probe was used. The 15-min Zeno SWATH DIA method used 85 variable windows over TOF MS 400–900 Da and MS/MS 140–1750 Da, Zeno trap pulsing on, 16 ms MS/MS accumulation, with an MS1 survey scan (400–1500 Da, 50 ms) before each cycle.
Download:
All raw files + FASTA (single archive) from the ProteoBench server
Also on ProteomeXchange PXD070049
FASTA (HYE mixed species + contaminants), contaminants from Frankenfield et al., JPR
Important
Do not rename the downloaded raw files.
How the metric is calculated#
For each precursor ion, ProteoBench sums the signal per raw file, removes contaminants (flagged
Cont_ in the FASTA) and precursors matching multiple species, log2-transforms the remaining
values, then computes the mean and coefficient of variation (CV) per condition. The difference
between the mean log2 intensity in A and B is compared against the expected log2 fold change for
that precursor’s species — that difference is epsilon.
The main plot shows the number of unique precursor ions quantified (vertical axis) against the mean or median absolute epsilon (horizontal axis).
Run your workflow#
Run your analysis on the downloaded files, then follow Your First Submission to upload, inspect, and (if you’d like) submit your results publicly.
Tool-specific setup#
Table: input files required for metric calculation and public submission
Tool |
Input file |
Parameter file |
|---|---|---|
AlphaDIA |
|
|
Custom |
|
— |
DIA-NN |
|
|
FragPipe |
|
|
FragPipe (DIA-NN quant) |
|
|
MSAID |
|
|
MaxQuant |
|
|
PEAKS |
|
|
Spectronaut |
|
|
Expand a tool below for setup details.
DIA-NN
Import raw files.
Add the FASTA, but do not select “Contaminants” (already included).
Turn on FASTA digest for library-free search / library generation.
Do not set verbosity/log level above 1, or parameter parsing will fail.
Upload
*_report.tsvor*_report.parquetfor scoring andreport.log.txtfor public submission.
AlphaDIA
Select the FASTA and import
.rawfiles in “Input files”.Define search parameters in “Method settings”.
Turn on “Predict Library”.
Turn on “Precursor Level LFQ”.
Required output files depend on version: AlphaDIA v1.x needs both
precursors.tsvandprecursor.matrix.tsv(or use the conversion notebook); later versions need onlyprecursors.parquet/.tsv.
Note
Version ≥1.10.4 is recommended for improved MS1-cycle checking.
FragPipe (DIA-NN quant)
Load the DIA_SpecLib_Quant workflow.
After importing raw files, assign experiments “by File Name”.
Do not add contaminants when adding decoys to the database.
Upload
*report.tsvfor scoring andfragpipe.workflowfor public submission.
FragPipe reports protein identifiers across two columns (“Proteins” and “Mapped Proteins”); these are concatenated to form the protein groups.
Spectronaut
Accepted format: BGS Factory Report — ..._Report.tsv for scoring,
..._Report.setup.txt for parameter parsing.
Import the module FASTA in “Databases” (UniProt parsing rule).
In “Analysis”, select “Set up a DirectDIA Analysis from folder” and load the raw-file folder.
Select the FASTA as the database.
Choose search settings.
Assign conditions:
LFQ_ZenoTOF8600_ZenoSWATH_85VW_15min_Nano_50ng_Condition_A_REP1–REP3to “A”, the equivalentCondition_Bfiles to “B”.Skip GO terms/library extensions.
Run the search, then export a BGS Factory Report as
..._Report.tsv.Upload
..._Report.tsvprivately and..._Report.setup.txtfor public submission.
MaxDIA (work in progress)
By default MaxDIA uses its own contaminants-only FASTA. This module’s FASTA already includes a
curated contaminant set, so untick “Include contaminants” (Global parameters → Sequences), and
set FASTA parsing to Identifier rule = >([^\t]*), Description rule = >(.*). Use “No Fractions”
and name experiments A_REP1–A_REP3, B_REP1–B_REP3.
Upload evidence.txt for scoring and mqpar.xml for public submission.
PEAKS
Rename samples (Sample 1→6) to match the .wiff file names exactly:
LFQ_ZenoTOF8600_ZenoSWATH_85VW_15min_Nano_50ng_Condition_A_REP1–REP3,
..._Condition_B_REP1–REP3.
ProteoBench automatically corrects small naming differences (case, spacing, minor typos) in PEAKS sample names, but anything it can’t confidently resolve will still fail with a message telling you which name(s) didn’t match. Exact names are still the safest bet.
Set Enzyme = trypsin, Instrument = ZenoTOF, Fragment = CID, Acquisition = DIA. In the workflow, use the Quantification option; define search parameters in “DB search”. In “Quantification” use “Label Free” (individually or grouped by condition); in “Report” set both Precursor/Peptide FDR and Protein Group FDR to 1%. Check “All Search Parameters” and the “Feature Vector CSV” (Export tab) once finished.
Custom format
If your tool isn’t listed above, upload a tab-delimited table with:
Sequence— unmodified peptide sequenceProteins—;-separated identifiers, including the species flag (e.g._YEAST)Charge— precursor chargeModified sequence— sequence with localized modifications, in ProFormaone quantitative column per sample:
LFQ_ZenoTOF8600_ZenoSWATH_85VW_15min_Nano_50ng_Condition_A_REP1…..._Condition_B_REP3
The table must not contain non-validated ions.
How ProteoBench maps each tool’s columns#
Each tool’s output format is described in a .toml file under
proteobench/io/parsing/io_parse_settings/, defining [mapper] (column-name mapping),
[condition_mapper]/[run_mapper] (raw file to condition/sample), [species_mapper] (protein-ID
suffix to species), [general] (contaminant flag Cont_ and decoy flag), and
[modifications_parser] where needed.
Result columns#
After upload, the results table includes: the precursor ion (modified sequence + charge); mean and standard deviation of log2-transformed and of raw intensity per condition; CV per condition; the difference of mean log2 values between conditions; per-raw-file intensity; the number of raw files with a non-missing value; species and whether the sequence is species-specific; the expected ratio for that species; and epsilon.
Use the slider to set the minimum number of raw files a precursor must be quantified in to be included in the plot.
Parameters tracked for public submission#
Upload your parameter file under “Download calculated ratios”; see Tool-specific setup above for which file that is per tool. ProteoBench tracks:
software tool name and version; search engine name and version, if different
FDR threshold (PSM, precursor, peptide, protein level)
match-between-runs (on/off)
precursor and fragment m/z range and mass tolerance
enzyme (Trypsin, for this dataset) and maximum missed cleavages
minimum/maximum peptide length
fixed and variable modifications, and the maximum number of modifications
minimum/maximum precursor charge
If any of these are missing from your parameter file, add them in the Comments for submission
field.
Check your parameter file for personal information
Parameter files can embed local file paths (FASTA location, .wiff/.wiff.scan data location, tool
installation paths) that may reveal personal or institutional information. Review and sanitize
before submission.