LFQ, human plasma#
beta Quantification — DIA, timsTOF, human plasma background (PYE)
This module is in beta stage
This module has been validated and is suitable for submissions. However, metrics or thresholds may be refined as we gather more community results and feedback.
Compares the sensitivity and quantification accuracy of workflows for data-independent acquisition (DIA) data on human plasma samples spiked with yeast and E. coli (the PYE dataset: Plasma/Yeast/E. coli).
At a glance#
Acquisition |
DIA |
Instrument |
timsTOF, nano-LC |
Level |
Precursor ion |
Metric |
Spike-in log2FC error, dynamic range, plasma accuracy |
What this module tests#
This module is best suited to evaluate the impact of, among others:
search engine identification
peak picking
low-level ion signal normalization
performance on complex biological matrices (plasma background)
It is not designed to evaluate later-stage post-processing of quantitative data (e.g. missing value replacement or manual filtering); please upload data without that kind of post-processing.
Dataset#
Based on the benchmark study by Distler et al., Nat. Comms., the PYE dataset uses a three-component mixture:
HUMAN (plasma background) — the complex endogenous proteome, expected ratio A/B = 1.0 (log2FC = 0)
YEAST (spike-in) — expected ratio A/B = 1/3 (log2FC ≈ −1.585)
ECOLI (spike-in) — expected ratio A/B = 2.0 (log2FC = 1)
Samples are commercial peptide digest standards (E. coli: Waters Corporation; S. cerevisiae:
Promega; human plasma as the complex background), mixed to the ratios above, with six technical
replicates per condition (12 raw files total). Acquisition used a timsTOF with nano-LC DIA. Raw
files follow the naming A9_G_DIA_nLC_tTOF_R1–R6 (condition A) and B9_G_DIA_nLC_tTOF_R1–R6
(condition B).
Download:
Or from ProteomeXchange JPST003358:
A9_G_DIA_nLC_tTOF_R1–R6.d,B9_G_DIA_nLC_tTOF_R1–R6.dFASTA (HYE mixed species + contaminants), contaminants from Frankenfield et al., JPR
Important
Do not rename the downloaded raw files.
How the metric is calculated#
Contaminant sequences (flagged Cont_ in the FASTA) and precursor ions matching multiple species,
or not quantified in any raw file, are removed. Quantities are log2-transformed, and the mean per
condition is computed with standard deviation and CV. For each precursor ion, the difference
between mean log2 in condition A and B is compared to its expected value (epsilon).
Main plot dimensions:
X-axis: absolute log2 fold-change error for spike-ins (YEAST + ECOLI), as Median or Mean
Y-axis: number of quantified spike-in precursor ions
Dot size: dynamic range of HUMAN plasma precursors (mean condition-wise log10 90th–10th percentile spread)
Dot opacity: HUMAN plasma quantification accuracy (absolute epsilon; darker = more accurate)
Calculation modes (available in Mean and Median variants):
Global: error calculated across all spike-in precursors together
Species-weighted: per-species error (YEAST, ECOLI), averaged equally
A cutoff slider filters precursors by the minimum number of runs in which they’re observed.
Dynamic range calculation#
The dynamic range summarises over how many orders of magnitude a workflow quantifies the human plasma background. It is calculated on the HUMAN precursor ions only, and for each condition separately:
Take the mean intensity of each HUMAN precursor ion in that condition (
Intensity_mean_AorIntensity_mean_Bin the intermediate table).Log10-transform these mean intensities.
Subtract the 10th percentile from the 90th percentile of the resulting distribution.
The values of condition A and condition B are then averaged:
dynamic_range = mean( P90(log10 I_A) - P10(log10 I_A),
P90(log10 I_B) - P10(log10 I_B) )
Because the values are log10-transformed, the dynamic range is expressed in orders of magnitude. A dynamic range of 3.0 means that the central 80% of the quantified plasma precursor ions span three orders of magnitude in intensity. A larger value indicates that the workflow quantifies both high-abundant and low-abundant plasma precursor ions, whereas a smaller value indicates that quantification is restricted to a narrower abundance window.
The 10th and 90th percentiles are used instead of the minimum and maximum so that the metric is not determined by single extreme precursor ions. The dynamic range is recomputed for every value of the cutoff slider, and the condition-wise values (dynamic_range_human_plasma_A and dynamic_range_human_plasma_B) are stored next to the averaged value in the submitted datapoint. The same explanation is available as a tooltip in the web interface: hover over the caption below the overview plot, or open the “How are the metrics calculated?” panel at the top of the page.
Identification counts#
The number of quantified precursor ions is reported both for the spike-ins combined and for each species separately:
Metric |
Description |
|---|---|
|
YEAST plus ECOLI precursor ions (the y-axis of the overview plot) |
|
HUMAN plasma background precursor ions |
|
YEAST spike-in precursor ions |
|
ECOLI spike-in precursor ions |
The per-species counts are stored for every cutoff level and are shown in the hover text of the overview plot as well as in the benchmark results table. They allow the identification depth in the complex plasma background to be assessed separately from the depth reached for the spike-in species, which is relevant because the two behave differently in a high dynamic range matrix.
In-depth plots#
For a single benchmark run the following plots are available in the “View Single Result” tab:
Abundance range plot: the normalised precursor intensity against the intensity rank, with the rolling median of the absolute epsilon on the secondary y-axis. A species is selected with the dropdown menu; the rank is assigned within the selected species, so the count axis runs from 1 to the number of precursor ions quantified for that species. The intensities remain normalised against the highest intensity over all species, so that the species can be compared on a common intensity scale.
Missing values plot: distribution of the percentage of missing values of the quantified HUMAN precursor ions against their intensity rank.
Log2 fold change distributions: density of the measured log2 fold changes per species, with the expected ratios indicated.
Coefficient of variation: CV distribution in condition A and B.
Signed epsilon plot: distribution of the signed epsilon (measured minus expected log2 fold change) per species, shown as a violin plot with an embedded box plot and a reference line at zero. Values below zero indicate that the log2 fold change is underestimated, values above zero that it is overestimated. The median signed epsilon and the percentage of precursor ions on either side of zero are annotated per species. A distribution centred on zero indicates an unbiased workflow. Note that ratio compression shifts the two spike-in species in opposite directions, because YEAST is expected below zero (log2FC = -1.585) and ECOLI above zero (log2FC = 1).
MA plot: measured log2 fold change against the mean abundance, coloured by species.
Run your workflow#
Run your analysis on the downloaded files with the tool of your choice; the module supports multiple input formats to maximize flexibility. Then follow Your First Submission to upload, inspect, and (if you’d like) submit your results publicly. The web app is where you upload results and submit them for public review.
Tool-specific setup#
Table: input files required for metric calculation and public submission
Tool |
Input file |
Parameter file |
|---|---|---|
AlphaDIA |
|
|
Custom |
|
— |
DIA-NN |
|
|
FragPipe (DIA-NN quant) |
|
|
PEAKS |
|
|
Spectronaut |
|
|
Expand a tool below for setup details.
DIA-NN
Generate a spectral library from the provided FASTA using DIA-NN’s library-generation mode.
Process raw files with the standard DIA-NN DIA workflow.
Export results as
*_report.tsvor*_report.parquet.Collect
*_report.log.txtfor public submission.
AlphaDIA
Process DIA raw files with AlphaDIA’s standard workflow.
Both
precursors.tsv(long-format precursor-level data) andprecursor.matrix.tsv(quantification matrix) are required and must both be uploaded.If your AlphaDIA version outputs a different format, preprocess with the conversion notebook.
Upload
log.txtfor public submission.
Spectronaut
Generate a spectral library from the provided FASTA.
Process raw files with DirectDIA or the standard Spectronaut DIA workflow.
Export a BGS Factory Report as
*_Report.tsv.Use
*_Report.setup.txtfor public submission.Ensure your export includes precursor-level data with modified peptide sequence, charge, protein IDs, and intensity columns.
FragPipe (DIA-NN quant)
Load the DIA_SpecLib_Quant workflow.
Import DIA raw files and assign experimental group information.
Generate or use a spectral library from the provided FASTA.
Do not add contaminants when adding decoys to the database.
Run the analysis and export the DIA-NN-style
*_report.tsv(precursor-level quantification).Upload
fragpipe.workflowfor public submission.
FragPipe reports protein identifiers across two columns (“Proteins” and “Mapped Proteins”); these are concatenated to form the protein groups.
PEAKS
Import DIA raw files into a new PEAKS project.
Configure sample grouping to match Condition A vs. Condition B.
Set up DIA quantification with appropriate parameters.
Use label-free quantification (LFQ) with Identification Directed Quantification (IDQ).
Export a text report (
.txt) with precursor-level quantification.Upload the settings
.txtfile for public submission.
Custom format
If your tool isn’t listed above, upload a tab-delimited table with:
Column |
Description |
|---|---|
|
Unmodified peptide sequence |
|
Sequence with localized modifications, ProForma format |
|
|
|
Precursor charge state |
|
Intensity values, condition A replicates |
|
Intensity values, condition B replicates |
The table should contain only validated ions, with no contaminant sequences or non-specific peptides.
Parameters tracked for public submission#
To submit publicly: upload your quantification output, the matching parameter/log file(s), fill in optional comments, confirm the metadata, and submit. A GitHub pull request is generated automatically for tracking and community review.
Contaminants are expected flagged Cont_; species annotation in protein identifiers must support
_HUMAN, _YEAST, and _ECOLI mapping.